GENERAL
1、Used in hospital, health center for fluoroscopy, radiography.
2、Rotation anode, full wave rectification.
3、Combined X-ray generator.
4、Equipped with real-time radiography device, motor-driven ray filter, light-locating beam limiter and motor-driven ray table.
5、Simple operation and safe, reliable in use.
Main Technical Parameter |
X-ray generator | Rotary anode, single focus, and bridge silicon rectification, The X-ray generator is an oil-immersed self-cooled unit. |
Max rated capacity | Fluoroscopy | 75kVp, 3mA |
Radiography | Current | 15mA | 30mA | 60mA | 100mA |
Voltage | 90kVp | 90kVp | 90kVp | 80kVp |
Time | 6.3s | 6.3s | 2.0s | 1.0s |
Power supply | Voltage | 180~240V |
Frequency | 50Hz |
Power | not lower than 7kVA |
Resistance | not larger than 1Ω |
Voltage adjustment range of x-ray tube | 45~90kVp (continuous) |
Range of the timer | 0.08~6.3 s |
Fluorescent screen size | 356 mm×356 mm |
Distance between the focus to the screen | 830 mm(max) 630 mm (min) |
Tube head carriage movement range | vertical: 1100 mm horizontal: 260 mm |
Specification of X-ray tube | Model XD55 rotation anode single focus 0.8mm |
Beam limiter | Field coverage at 100 cm FFD, 45×45 cm max, 0×0 cm min |
Movement of the tabletop | 400 mm |
Field size of spot film device | 127mm×178mm (one piece) 203mm×254mm(one piece) 1/2(127mm×178 mm)(two pieces) |
Transport dimension, Weight | 2310×1000×1250 (Net)418/(Gross)564 |