Maximum infusion rate | 500ml/h (Different syringes have different maximum rate). |
Range of infusion rate | 10ml syringe: 0.1-100ml/h 20ml syringe : 0.1-200ml/h 50ml syringe : 0.1-500ml/h |
Bolus rate | 10ml syringe : 100ml/h 20ml syringe : 200ml/h 50ml syringe : 500ml/h |
Accumulated infusion volume | 0.1-999.9ml |
Accuracy | ±3% |
Power supply | AC110-240V, 50/60Hz, 25W. |
Battery | Rechargeable lithium battery, 7.4 V, 1650mAh. |
Maximum power consumption | 25W, running more than 5 hours after being fully charged. |
Battery charge | When syringe pump is connected to AC power, the power needs to be turned on (you may or may not operation the machine), the battery will automatically start charging (need8~14hrs). |
Displayed information | Flow rate, accumulated infusion volume, syringe size, battery capacity, bed No., AC power connecting indication |
Alarm information | infusion completion, near completion, occlusion, low battery, syringe abnormal, control abnormal, no AC alarm |
Maximum size of the shell | 288×130×122 mm(length×width×height) |
Maximum weight | <2.4kgs |
Classification | Type IIB |
Electrical Shock Proof | Battery: Class II; AC Power: Type BF. |
Shell material | ABS plastic |
Operation environment | Environment temperature +5℃--+40℃, atmosphere pressure 86~106kPa,relative humidity 20%~90% |
Storage and transport condition | Environment temperature -15℃- +50℃, atmosphere pressure 86~106kPa,relative humidity20%~90% |
Applicable syringes | 10ml, 20ml and 50ml syringe (use ‘Double Dove’ syringe to test). |
Operating mode | Continuous operation |