Features
1.This product meets the national standard GB9706.29 - 2006 requirements;
2.A full range of anesthesia breathing process control and monitoring, collection anesthesia, respiratory management, respiratory monitoring in one;
3.Breathing mode: IPPV, MANU, for adults and children;
4.High-definition LCD display, showing a variety of respiratory parameters and respiratory waveform;
5.Standardization of ventilation systems, airtight, easy to use, easy to clean and disinfect;
6.Classic ascending bellows, help you to judge quickly and accurately to ensure patient safety;
7.High-precision anesthetic evaporator, with the flow, temperature and pressure automatic compensation;
8.According to need any matching built-in oxygen concentration of carbon dioxide monitors and independent monitor;
9.Designed according to ISO9703 intelligent alarm system, can easily distinguish between primary and secondary respectively, in order to better patient care.
Anesthesia Machine
1. gas supply: pipeline gas can be used, or the central cylinder gas supply, rated working pressure 0.4MPa, permitted fluctuation range: 0.32 ~ 0.60MPa.
2, flow: single-tube or double tube flowmeter with oxygen / nitrous oxide and oxygen supply shortage of nitrous oxide automatic interlock automatic cut-off device to ensure the oxygen concentration of not less than 25%.
3, low-oxygen alarm and automatically cut off the source: oxygen gas source when the pressure is too low is greater than 7 seconds pneumatic audible alarm and automatically cut off the supply of nitrous oxide.
4, fast oxygen flow rate: (35 ~ 75) L / min.